Skip to main content

Table 3 Cyscope® FM results compared to LM results with respect to clinical status

From: Comparison of the fluorescence microscopy Cyscope® with light microscopy for malaria diagnosis in a small and active surveillance in Cameroon

 

Cyscope® FM

All

Febrile

Afebrile

Positive

Negative

Total

Positive

Negative

Total

Positive

Negative

Total

LM

Positive

63

12

75

4

3

7

59

9

68

 

Negative

110

256

366

17

32

49

93

224

317

 

Total

173

268

441a

21

35

56

152

233

385

  1. FM Fluorescence microscopy, LM Light microscopy
  2. Data are presented as frequency