Table 4 Diagnostic performances of Cyscope® FM with regard to clinical status

From: Comparison of the fluorescence microscopy Cyscope® with light microscopy for malaria diagnosis in a small and active surveillance in Cameroon

Parameters

All (n = 441)

Febrile (n = 56)

Afebrile (n = 385)

P value

Value

95% CI

Value

95% CI

Value

95% CI

Se (%)

84.00

74.08–90.60

57.14

25.04–84.18

86.76

76.71–92.87

0.74

Sp (%)

69.95

65.07–74.42

65.3

51.32–77.08

70.66

65.42–75.40

0.92

PPV (%)

63.58

56.19–70.38

19.05

7.67–40.0

38.81

31.44–46.75

0.3

NPV (%)

95.55

88.31–94.72

91.43

77.62–97.04

96.13

92.83–97.96

1

PLR

2.8

1.65

NA

2.95

NA

NLR

0.23

NA

NA

NA

NA

Accuracy (%)

89.36

83.34–90.21

10.08

51.2–75.85

79.27

68.89–77.67

0.63

Reliability (%)

53.95

22.45

57.43

Kappa

0.45

-

  1. CI Confidence interval, FM Fluorescence microscopy, Se Sensitivity, Sp Specificity, PPV Positive predictive value, NPV Negative predictive value, PLR Positive likelihood ratio, NLR Negative likelihood ratio, NA Not applicable. Pearson’s chi-square test was used to make comparisons. Significance was set at p value < 0.05